High speed handpiece maintenance procedures and precautions maintenance processes

1.clean water spraying holes: three holes clean water pipes and clean the probe with a three-hole jet,water jet clogging prevention;
2.Clean internal phone: Cleaners aligned phone tail,gently pressing 1 seconds,remove the dirt inside the phone;
3.the cleaning surface of the phone: the cleaner sprayed on a clean tissue,then gently wipe the surface of the phone;
4.lubrication head: There are two methods to choose from,recommend the first 1) plus grease: grease installed in accordance with the Canadian way of loading needle to the nose,then rotate refueling; please note that the new phone is rotated 360 degrees Old phone rotates 180 degrees,the use of lubricants,lubricating oil aligning phone tail pressing 1 seconds,then wipe with a clean paper towel.
5.the rotary bur: After refueling finished,the bur Replace the nose,the low speed of rotation 10-15 seconds; this allows the lubricant sufficiently uniform smear on the head inside; note that this work can also be After autoclaving conduct.
6.autoclavable: After the phone in the refueling autoclavable cabinet for disinfection,sterilization at 136 degrees for 10-30 minutes 2 atmospheres;
7.if there is no rotation of the bur After refueling,it is possible at this time to drive the needle Replace the phone at a low speed of rotation 10-15 seconds.



High speed handpiece maintenance:
1.refueling work must be done before autoclavable and not prior to use;
2.prohibit idling; easy to damage the collet;
3.try to use the latest clean lubricating product is not recommended the use of domestic products.If you use chlorine,hypochlorous acid and aqueous cleaning within the domestic lubricant,it will cause damage to the phone;
4.preventing the bump,or likely to influence the clamping device and bearing life,even the life of the phone;
5.the use of non-diameter requirements will cause bur clamping device failure (standard bur diameter 1.59-1.60 mm);
6.gas pressure and gas quality must comply with the requirements of the appropriate configuration for each phone.

Product details,please click: http://www.medicalequipment-msl.com/htm/dental-instrument/